Improving central line needle insertions using in-situ vascular reconstructions

Published in Computer Methods in Biomechanics and Biomedical Engineering: Imaging & Visualization, 2020

Recommended citation: Groves LA, Li N, VanBerlo B, Veinberg N, Peters TM, Chen ECS, (2021). "Improving central line needle insertions using in-situ vascular reconstructions"; in Computer Methods in Biomechanics and Biomedical Engineering: Imaging & Visualization, (9)4, pp. 363-369. https://doi.org/10.1080/21681163.2020.1835542

We developed a neck central line insertion guidance system that renders 3D ultrasound (US) surface reconstructions of the carotid artery (CA) and internal jugular vein (IJV), a tracked model of the needle, and needle trajectory on a 2D monitor. Twenty clinicians evaluated this system compared to US-only guidance on a phantom using time and insertion accuracy metrics. The 3D system had a 100% success rate compared to 70% for the US-only system. The average distance from the centre line of the US reconstructed IJV was 1.8±0.9 mm under 3D guidance compared to 4.2±2.9 mm for US-only. Our system significantly improved needle insertion success rates and targeting accuracy compared to the US-only approach through a radiation-free surface reconstruction of the neck vascular structures. This work has the potential to provide a mobile 3D imaging and visualisation system for needle-based vascular interventions.

Download paper here BibTeX